紫外分光光度法对二氧化氯含量的测定 |
|
王岙 王萍 朴红莲 马莹 王巍
提要 采用紫外分光光度法,于二氧化氯有特征吸收峰的260 nm波长下测定溶液中二氧化氯含量。对二氧化氯含量≤300 mg/L的溶液,测定结果的相对标准偏差≤5.09%。样品溶液pH值在5~10与放置时间≤50 min时,对测定结果影响不大。 关键词 二氧化氯 浓度测定 紫外分光光度法
MEASUREMENT OF CHLORINE DIOXIDE CONTENT BY ULTRAVIOLET SPECTROPHTOMETRY
Wang Ao Wang Ping Pu Honglian Ma Ying Wang Wei (Jilin Provincial Sanitary and Anti-epidemic Station, Changchun 130021)
Abstract
Ultraviolet spectrophotometry was used to measure chlorine dioxide content in solution at wavelength of 260 nm which is the characteristic absorption peak of chlorine dioxide. The relative standard deviation of measured results of the solution containing chlorine dioxide ≤300 mg/L was ≤5.09%. When the pH value of the sample solution was 5~10 and the solution was laid aside for ≤5 min, there was no significant influence on measured results. Key words chlorine dioxide measurement of concentration ultraviolet spectrophotometry
对消毒剂二氧化氯含量的测定,有报道用分光光度法[1~3]。该法只适用于测定低含量二氧化氯。我们采用紫外分光光度法测定,并报告结果于下。
1 方法
1.1 二氧化氯标准溶液的制备 向亚氯酸钠溶液中加入稀硫酸,使生成的二氧化氯气体溶入水中,得到二氧化氯溶液。标定溶液中二氧化氯含量后,稀释成二氧化氯含量为500 mg/L的标准溶液。 1.2 标准工作曲线的制备 取6支干燥的10 ml具塞比色管,除第1支外,向第2~6支比色管分别加标准溶液0.5、1.0、2.0、4.0、6.0 ml。然后,各管均加蒸馏水至刻度,盖塞摇匀。各管二氧化氯含量分别为0、25、50、100、200、300 mg/L。以蒸馏水(第1管)作参比,在紫外分光光度计的260 nm波长(该二氧化氯在此波长下有一特征吸收峰)处测其吸光度值。以二氧化氯含量与其对应吸光度值作图,得标准工作曲线。 1.3 样品二氧化氯含量的测定 取消毒液或其稀释液于260 nm波长处测出吸光度值。从标准工作曲线上查得该吸光度值对应的二氧化氯含量;若为稀释液,据稀释度换算出原液二氧化氯含量。
2 结果
2.1 二氧化氯溶液条件 2.1.1 pH值 将二氧化氯标准溶液稀释成含量为50 mg/L的溶液,调节成不同pH值,测其吸光度值。结果,在pH值为5~10之间变化,吸光度值变化不大,但以pH值为7时吸光度值较大(表1)。
表1 50 mg/L二氧化氯溶液pH值与吸光度值的关系
pH值 吸光度值(A) 5 0.226 6 0.220 7 0.233 9 0.224 10 0.223
2.1.2 溶液放置时间 将二氧化氯标准溶液稀释成含量为100 mg/L的溶液,放崐置不同时间,测其吸光度值。结果,放置时间在50 min内,吸光度值较稳定(表2)。为此,制作标准工作曲线时,不调节稀释液pH,但在配制后50 min内测吸光度值。
[1] [2] 下一页 上一个医学论文: 次氯酸钠复方消毒剂杀菌效果及毒性的试验观察 下一个医学论文: 成人肠套叠35例临床分析
|
|
|
|
|
|
|